A Radiation Levels Measurement at Workplace of Cyclotron-PET/CT Facilities (in Karachi) using TLD Dosimeters

Authors

  • Kauser Perveen PNRA
  • Ayesha Mohsin Pakistan Nuclear Regulatory Authority, Islamabad, Pakistan
  • Hira Nadeem
  • Syed Mishkat Ali
  • Khalid Hussain Pakistan Nuclear Regulatory Authority, Islamabad, Pakistan

DOI:

https://doi.org/10.33317/ssurj.385

Keywords:

Radiation Levels, Cyclotron-PET/CT, TLD

Abstract

Cyclotron is the most widely used proton particle accelerator for production of short lived radionuclides particularly F-18 for diagnostic purpose in oncology imaging. Currently, only Fluorine [F-18] is being produced in Pakistan. F-18 is tagged with glucose analogue and converted into [18F] FDG (Fluorodeoxyglucose) by synthesis process. [18F] FDG emits positron and, via annihilation process, produces two gamma photons of 511keV energy, which is of great concern for occupational safety. Therefore, a study was carried out to assess the dose incurred by radiation workers working in controlled and supervised areas of cyclotron facility.

In this study, the ten (10) calibrated Thermoluminescene dosimeters (TLD-100) were annealed and placed in different locations of four Cyclotron-PET/CT facilities for determination of dose in work area. In the selected cyclotron facilities, one out of four was unshielded cyclotron and other three were self-shielded. Ten (10) TLDs were placed in preselected controlled and supervised areas of facilities for the duration of 30 days. After four weeks of exposure, TLDs were removed and assessed on TLD reader for dose evaluation.

The evaluated results of radiation levels from TLDs are in good agreement with the criteria for classification of areas as defined in Regulations on Radiation Protection (PAK/904). The measured doses of work area are well below the defined regulatory dose limits for controlled area (>6mSv/yr) as well as supervised area (>1mSv/yr) except in case of unshielded cyclotron vault. It is concluded that the facilities have implemented the requirement 27 of PNRA Regulations on Radiation Protection (PAK/904) and also have established strong radiation safety measures for minimization of occupational exposure.

References

IAEA. (2008). Cyclotron Produced Radionuclides: Principles and Practice (465). Technical Report Series. Retrieved from: https://www.iaea.org/publications/7849/cyclotron-produced radionuclides-principles-and-practice

IAEA. (2009). Cyclotron Produced Radionuclides-Physical Characteristics And Production Methods (468). Technical Report Series. Retrieved from:https://www.pub.iaea.org/mtcd/publications/pdf/trs468_web.pdf

Schmor, P. W. (2010). Review of Cyclotrons Used in the Production of Radio-Radioisotopes for Biomedical Applications. Retrieved from:https://accelconf.web.cern.ch/cyclotrons2010/papers/frm2cio01.pdf

Pakistan Nuclear Regulatory Authority. (2019). Regulations on Radiation Protection (PAK/904) Rev.01, Pakistan. Retrieved from: https://www.pnra.org/regulations.html

Zargan, S., Ghafarian, P., Monfared, A. S., Sharafi, A. A., Bakhshayeshkaram, M., & Ay, M. R. (2017). Evaluation of radiation exposure to staff and environment dose from [18F]-FDG in PET/CT and cyclotron center using thermoluminescent dosimetry. Journal of biomedical physics & engineering, 7(1), 1.

Biegała, M., & Jakubowska, T. (2019). Levels of exposure to ionizing radiation among the personnel engaged in cyclotron operation and the personnel engaged in the production of radiopharmaceuticals, based on radiation monitoring system. Radiation Protection Dosimetry, 189(1), 56-62. DOI: https://doi.org/10.1093/rpd/ncaa012

NCRP, N. (2005). Report 144: Radiation Protection for Particle Accelerator Facilities. In National Council on Radiation Protection and Measurements.

Avila, O., Torres-Ulloa, C. L., Medina, L. A., Trujillo-Zamudio, F. E., Gamboa-deBuen, I., Buenfil, A. E., & Brandan, M. E. (2011). TL measurement of ambient dose at a nuclear medicine department. Radiation Measurements, 46(12), 1843-1846. DOI: https://doi.org/10.1016/j.radmeas.2011.06.007

Kaur, A., Singh, B., & Mittal, B. R. (2012). Radiation surveillance in and around cyclotron. Indian Journal of Nuclear Medicine, 25(3), 74. DOI: https://doi.org/10.4103/0972-3919.115395

Marwa, A. A., Naima, A. B., & Bailey, D. (2019). Estimation of radiation exposure to workers during [18F] FDG PET/CT procedures at molecular imaging center, Oman. Journal of medical imaging and radiation sciences, 50(4), 565-570. DOI: https://doi.org/10.1016/j.jmir.2019.05.009

IAEA. (2018). Occupational Radiation Protection-IAEA Safety Standards (GSG-7). General Safety Guide. Retrieved from: https://www-pub.iaea.org/mtcd/publications/pdf/pub1785_web.pdf

McKinlay, A. F. (1981). Thermoluminescence dosimetry (medical physics handbooks 5). Adam Hilger.

Cameron, J. R., Suntharalingam, N., & Kenney, G. N. (1968). Thermoluminescent Dosimetry. The University of Wisconsin Press, madison, Milwaukee, and London.

Santana, P. D. C., Oliveira, P. M. C. D., Mamede, M., Silveira, M. D. C., Aguiar, P., Real, R. V., & Silva, T. A. D. (2015). Ambient radiation levels in positron emission tomography/computed tomography (PET/CT) imaging center. Radiologia Brasileira, 48, 21-25. DOI: https://doi.org/10.1590/0100-3984.2013.1821

Guillet, B., Quentin, P., Waultier, S., Bourrelly, M., Pisano, P., & Mundler, O. (2005). Technologist radiation exposure in routine clinical practice with 18F-FDG PET. Journal of nuclear medicine technology, 33(3), 175-179.

Harshaw, T. L. D. Model 6600 Plus Automatic TLD reader user manual (2009). Thermo Fisher Scientific Inc., Retrieved from: http://www. kortec. cz/files/thermo/TLDModel6600. pdf.

IAEA. (2014). Radiation Protection and Safety of Radiation Sources: International Basic Safety Standards . General Safety Requirements Part 3. Retrieved from: https://www.pub.iaea.org/mtcd/publications/pdf/pub1578_web-57265295.pdf

Downloads

Published

2023-06-28

How to Cite

Perveen, K., Mohsin, A., Nadeem, H., Ali, S. M., & Hussain, K. (2023). A Radiation Levels Measurement at Workplace of Cyclotron-PET/CT Facilities (in Karachi) using TLD Dosimeters. Sir Syed University Research Journal of Engineering & Technology, 13(1), 35–41. https://doi.org/10.33317/ssurj.385